Lucene search

K
ContechealthCms8000 Firmware

5 matches found

CVE
CVE
added 2022/09/13 3:15 p.m.44 views

CVE-2022-38069

Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device para...

6.1CVSS5.4AI score0.00066EPSS
CVE
CVE
added 2022/09/13 3:15 p.m.41 views

CVE-2022-38100

The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent th...

7.5CVSS7.3AI score0.00106EPSS
CVE
CVE
added 2022/09/13 3:15 p.m.40 views

CVE-2022-3027

The CMS8000 device does not properly control or sanitize the SSID name of a new Wi-Fi access point. A threat actor could create an SSID with a malicious name, including non-standard characters that, when the device attempts connecting to the malicious SSID, the device can be exploited to write arbi...

5.7CVSS6AI score0.00019EPSS
CVE
CVE
added 2022/09/13 3:15 p.m.39 views

CVE-2022-36385

A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive...

6.8CVSS6.9AI score0.0008EPSS
CVE
CVE
added 2022/09/13 3:15 p.m.35 views

CVE-2022-38453

Multiple binary application files on the CMS8000 device are compiled with 'not stripped' and 'debug_info' compilation settings. These compiler settings greatly decrease the level of effort for a threat actor to reverse engineer sensitive code and identify additional vulnerabilities.